The ultimate solution in wireless thermal validation.

Ellab has incorporated the latest in electronic technology and innovation to deliver a wireless multi-channel data logger unmatched in accuracy, performance and versatility. The flexible system is easily adapted to your requirements and provides an elegant solution to the most demanding validation applications.

 
Open online brochure
 

Accuracy
The incorporation of state of the art technology and extensive testing have resulted in:

• Temperature: ±0.05 °C
• Humidity: ±2%
• Pressure: ±0.25% full scale
• Time: ±5 sec. per 24 hours
• Rotation: ±1%

Compare and you will find TrackSense® Pro the most accurate data logger available on the market today.

Performance
The robust loggers are designed to operate under extreme conditions without ever loosing valuable data. They can tolerate temperatures from -80 °C to +150 °C and withstand pressures up to 10 Bar fully immersed. The non volatile memory stores up to 30,000 data points. Recent improvements in battery technology have dramatically improved performance and lifetime even at high temperatures.

Versatility
Over 100 sensor configurations are available in stock with approximately 30% of our production dedicated to developing custom designs for specific applications. Whatever the package, container, tunnel or vessel Ellab has a sensor appropriate for your application. If not, we will design it. All of these sensors whether rigid, flexible, multi-channel, temperature, humidity or pressure are interchangeable on the TrackSense® Pro Logger.

The small size and convenience of wireless data loggers make them exceptionally easy to use in countless validation applications such as;

 

      Pharmaceutical: 

      Food:

  • Retorts
  • Pasteurization 
  • Spiral Cooker/Coolers
  • Hydrostats
  • Conveyorized Ovens 
  • Spiral Freezers 
  • Blast Freezers 
  • Cold Storage Chain
  • Smoke Houses
Features: Benefits:

Wireless System

No cables in the process.
Less time due to easy setup, even in continuous processes

16 Loggers Reader Station

Automated multiple download saves time

Small Size Logger and 2 mm Sensor

Minimal Mass to effect Process

High Accuracy

Most accurate and reliable measurements of any wireless system

Multiple Point User Calibration

Flexible and reliable at the same time

Large Non-volatile Memory

30,000 data points per logger.
Huge range of applications. Long-term monitoring possible

Field Replaceable Battery

1,000 hours of operation

System Security

Encrypted files, Electronic signatures, and Audit trail. Compliant to 21 CFR part 11

Interchangeable Sensors

Greater flexibility for temperature, pressure and humidity measurements

Quad Sensors

4 channels on one logger.
Saving money and space. Minimal effect on process

Customised Validation Reports

Ability to summarise and report key data as preferred

 

Detailed Control of Validation Studies
The ValSuite Software documents and guides you through the complete thermal validation process. The database structure in the software enables complete documentation and procedural control for the operators.

Test Setup
Templates allow detailed test criteria to be pre set in the software by the assigned administrator. Information on sensor placement, operator, test, vessel, required temperature limits, start and stop time, monitoring interval and specific calculations can all be pre set and saved. This ensures accurate documentation and correct implementation of required procedures of consistent repeatable tests.

 

Pre identifying sensor positions greatly reduces setup time and errors. A test setup sheet is indication where each sensor should be placed. Later during analysis this information is easily displayed and tracked.

Software Data Analysis Features
Data analysis tools greatly reduces the time needed to find critical data. The ability to zoom graphically and display up to 8 tiled windows at once simplifies identifying import data. Multiple calculations such as min/max standard deviation and average, deltaT and lethality can be calculated on any block of data displayed eliminating the need to export data and impoving data security.

ValSuite collects and presents validation data from both realtime E-Val Flex and TrackSense Pro wireless data logging systems. The data from both systems can be presented and analyzed in the same session. The system can run up to 128 channels which can be identified and displayed in different groups such as penetration and distribution. Any grouping or specific channels can be displayed in a separate data block and analyzed.

 

Outputting Reports
A complete thermal validation report can be produced with pass/fail criteria, detail on mapping positions, operator and vessel ID, calibration offsets for sensors, real data and statistical summaries on the data. ValSuite also maintains templates for reports desigend to meet specific requirements of tests such as EN554. The templates can be customized to organize the data and perform calculations to your exact criteria. this saves tremendous time teducing the data alalysis process. Reports can be reviewed with the print preview feature and output in a pdf file document.

 
Maintain Security
ValSuiteTM Pro operating software is designed to maintain the highest levet of security for compliance to FDA 21 CFR Part 11. The software administrator controls and assigns user ID and access privileges. Any user function can be deactivated, removing the software feature from view and access of specified users. This reduces the risk of mistakes significantly giving tight control of usage by the administrator.
 

Compliant to FDA Guidelines
•     Raw data is built into a database structure and cannot be modified or erased.
•     Complete audit trail with easy search engine to find data by date, session, user and module.
•     User ID and passwords with detailed user access control.
•     Sensor ID provides complete NIST traceability.
•     Customized report generator eliminating need to export data.
•     Specific compliance reports (e.g. EN554).
•     Automated pre and post calibration reports.
•     Print preview feature with report output in pdf file format.

GAMP4 guidelines and DS/EN ISO 9001
All documentation for development of ValSuiteTM Pro software is in accordance with the guidelines set out in GAMP4. Software package includes appropriate documentation. Ellab quality system is compliant with DS/EN ISO 9001.

Validation Report and Regulatory Compliance
Three regulatory compliance documents are available. Ellab supplies with the software a complete Validation Guide, Validation Report, and