You are here:» Products» E-Val Flex» Specifications
REPLACEMENT OF
Mercury-in-glass
 

E-Val Flex Modules

 

The E-Val Flex modules are available with 8 or 16 type T thermocouple channels and the master modules have additional channels for digital pressure and deflection.

Non-volatile memory in all modules stores data that has not been transferred to the PC in case of power failure or loss of connection to PC. The 8 MB memory is sufficient for a minimum of 8 hours using 16 channels with sampling every second for each channel.

The RS485 port on the master module allows pressure measurements. Multiple pressure sensors can be connected to the same RS485 port.

The sampling rate can be set from 1 second to 24 hours and is independent from the number of channels.
 

 
Technical Specifications

Operating Range: From 5°C to 50°C

 

 
Temperature  
Measuring Range: -200°C to +400°C
Measuring Accuracy at 23°C :

±0.05°C

Total System Accuracy at 23°C:

±0.1°C

Resolution:

0.01°C

   
Pressure  
Measuring Range:                                          0 to 4 Bar (abs.)
Measuring Accuracy:                                             ±0.15% FS
Resolution:                                                1 mBar
   
Deflection  
Measuring Range:                                             -8 to +8mm   
Resolution:                                               0.01mm
   
Sample Rate: 

1 Second to 24 Hours

Weight:

    1.9 kg

Power Consumption:                                        Less than 5 Watts

 
« Back