jpgFeature/sdesk 696 implement raw brick http

Oxygenator Probes

The probes used in oxygenators are offered in several designs to fit the different equipment on the market. Three probe types are described in the technical specifications below, contact Ellab for custom sizes.

 

Contact Sales

Technical Specifications

Design:

Type MIC

Thermocouple Type:

Cu/CuNi

Diameter:

2.4 mm

Length:

35 mm

Material of Electrode:

Stainless Steel

Response Time:

1.1 Sec.

Accuracy in the Range +15 to +45 °C:

±0.1 °C

Cable Length:

2.75 m

Design:

Type MIA

Thermocouple Type:

Cu/CuNi

Diameter:

3.0 mm

Length:

30 mm

Material of Electrode:

Stainless Steel

Response Time:

2.1 Sec.

Accuracy in the Range +15 to +45 °C:

±0.1 °C

Cable Length:

2.75 m

Design:

Type MIE

Thermocouple Type:

Cu/CuNi

Diameter:

3.8 mm

Length:

25 mm

Material of Electrode:

Stainless Steel

Response Time:

3.7 Sec.

Accuracy in the Range +15 to +45 °C:

±0.1 °C

Cable Length:

2.75 m

 

Patient Safety:

Desinfection:
Sterilization:

 

The Probes have passed an isolation test of 0.5kV.
The Silicone Cable is inactive and does not irritate the skin
70% alcohol, 0.5% chlorexidine and 29.5% water
Autoclaved up to 120 °C or gas sterilized by EtO