jpgFeature/sdesk 696 implement raw brick http

Surface Probes

The surface temperature probes are mainly used for measuring temperature of skin, fingers and toes.
There are six types:

1. Type MHA is for instant measuring within a narrow area all over the body.
2. Type MHB is for instant and stationary measuring of peripheral temperature which is fixed with tape.
3. Type MHC is for stationary measuring of peripheral temperature. This is to be placed under the nail of fingers and toes and fixed with tape.
4. Type MHD is for instant and stationary skin measuring all over the body.
5. Type MHE is for stationary measuring of skin, toes and fingers of babies (especially incubator babies).
6. Type MHF is for stationary measuring of skin, fingers and toes of adults.

 

Contact Sales

Technical Specifications

1. MHA

Thermocouple Type:

Cu/CuNi

Size of Contact Point:

7.0 mm

Size of thickness:

0.9 mm

Material of Electrode:

Stainless Steel Tube

Response Time:

0.15 Sec.

Accuracy in the Range +15 to +45 °C:

±0.1 °C

Cable Length:

2.75 m

2. MHB

Thermocouple Type:

Cu/CuNi

Size of Contact Point:

8.0 mm

Size of thickness:

0.9 mm

Material of Electrode:

Stainless Steel Tube

Response Time:

0.15 Sec.

Accuracy in the Range +15 to +45 °C:

±0.1 °C

Cable Length:

2.75 m

3. MHC

Thermocouple Type:

Cu/CuNi

Size of Contact Point:

40 mm

Size of thickness:

1 mm

Material of Electrode:

FEP Tube

Response Time:

0.9 Sec.

Accuracy in the Range +15 to +45 °C:

±0.1 °C

Cable Length:

2.75 m

4. MHD

Thermocouple Type:

Cu/CuNi

Size of Contact Point:

0.1 mm

Size of thickness:

0.1 mm

Material of Electrode:

Flexible Plast Foil

Response Time:

0.15 Sec.

Accuracy in the Range +15 to +45 °C:

±0.1 °C

Cable Length:

2.75 m

5. MHE

Thermocouple Type:

Cu/CuNi

Size of Contact Point:

5.0 mm

Size of thickness:

2.2 mm

Material of Electrode:

Polyoxymethylen

Response Time:

2.2 Sec.

Accuracy in the Range +15 to +45 °C:

±0.1 °C

Cable Length:

2.75 m

6. MHF

Thermocouple Type:

Cu/CuNi

Size of Contact Point:

18.0 mm

Size of thickness:

5.8 mm

Material of Electrode:

Stainless Steel Plate

Response Time:

13.5 Sec.

Accuracy in the Range +15 to +45 °C:

±0.1 °C

Cable Length:

2.75 m

Follow Us @ Linkedin
Join Us @ Youtube
Contact Us Today
Newsletter Sign up
Sign Up Now

This website uses cookies to improve your experience. We'll assume you're ok with that, but you can opt-out if you wish.

Read more.
Close