No results
Card image cap
feature icon

Temperature Range of
-196 to +420 °C

Card image cap

Ellab Temperature Standard

ETS25

The traceable ETS allows for extremely accurate temperature measurements by converting resistance into temperature using its built-in algorithms.

When connected to the ValSuite software, the LiquiCal bath and ETS provide a perfect setup which is automatic. This saves time and allows for a convenient calibration.

Other Options

Key Features and Benefits

checkmark

Temperature Range -196 to +420 °C

checkmark

Accuracy ±0.025 °C

checkmark

Separated electronics from heat zone

checkmark

Robust and Versatile Design

ETS25

Temperature Range:

-196 to +420 °C

Accuracy – Includes Instrument, PRT and Drift:

-50 to +150 °C:

±0.025 °C

Other temperatures:

±0.025 °C

Measuring Sample Rate:

4 Sec.

Digital Filter:

0-10 Sec.

Resistance:

100 Ω

Sensor Response Time:

T-63%:

33 Seconds

T-90%:

45 Seconds

Overall Dimension:

ø 16 x 475 mm

Interface & Power Source:

USB

Sensor Material:

ANSI 316L

Dimension of Sensor:

ø 6.35 x 305 mm

Minimum Immersion Depth:

100 mm

Environmental:

15 – 35 °C / 0 – 90% RH, non-condensing

Calibration:

Traceable to NPL

Calibration Algoritms:

ITS-90

Recommended Calibration Interval:

12 months

Weight w / wo Display:

0.35 kg / 0.25 kg

Optional Display

Display:

8 digits, 7 segments, green LED

Resolution:

0.001°C

Power Supply for “Stand-Alone” use:

9 Volt

software image