WIRED THERMOCOUPLE SYSTEM TECHNICAL SPECIFICATIONS THERMOCOUPLES PRESSURE & ADDITIONAL PARAMETERS
WIRED THERMOCOUPLE SYSTEM
E-Val™ Pro
The E-Val Pro thermal validation system allows for easier access to controlled environments due to its compact size and light weight. The flexible configuration can handle up to 40 thermocouples per module. The large internal memory is capable of holding up to 10 studies.

The E-Val Pro also contains a rechargeable internal battery that can power the module for 8 hours. The validator can use the E-Val Pro as a stand-alone unit or it can be connected via LAN/USB.

E-Val Pro utilizes ValSuite™ Pro validation software with powerful reporting tools. This allows for validators to combine data loggers and thermocouples in the same study. Software and hardware are developed according to GAMP and comply with requirements from FDA and other regulatory bodies.

 

Interchangeable Sensor Array

  • Can be removed for service and calibration
  • Contains precision voltmeter
  • 4 channel multipurpose, accommodates: TC type T and K (ready for type J, B, E, N, R and S), 4-20 mA, 0-10 V and I/O relay
  • 12 channel TC type T and K (ready for type J, B, E, N, R and S) as well as other low power analog/digital inputs
     
Contact Sales
TECHNICAL SPECIFICATIONS
TECHNICAL SPECIFICATIONS

Operating Range:

5°C to 50°C / 41°F to 122°F

Environmental: 

0 - 90% RH, non-condensing

Sample Rate:

1 Sec. to 24 Hours

Weight:

3.0 kg (including battery and 40 channels)

Electrical Protection Class:

IP40

Power Consumption:

Less than 9 Watts (40 temperature sensors)

Memory:

40 channels 1 sec. sample rate in 8 hours for 10 sessions

Number of channels:

40 channels

Time:

Time Accuracy:

±1 Sec. per 24 Hours

Temperature:

Measuring Range: 

-200 to +1.300 °C (ready for -270 °C to +1.820 °C) depending on sensor type

Measuring Range (Type T):

-200 to +400 °C

Measuring Accuracy at an ambient temperature of 23 °C:

±0.05 °C between -100 and +400 °C / ±0.1 °C between -200 and <-100 °C

Total System Accuracy at an ambient temperature of 23 °C:

±0.10 °C between -50 and +150 °C (type T)

Resolution:

0.01 °C

Pressure:

Measuring Range:

0 to 4 Bar ABS

Measuring Accuracy:

±0.15% FS

Resolution:

1 mBar

Other Parameters:

Measuring Range:

4-20mA or 0-10V

THERMOCOUPLES
Thermocouple Sensors

Interchangeable sensor arrays are supplied and each can accommodate up to 12 thermocouple or analog/digital inputs. Thermocouples can be organized into groups for real time statistical analysis.

The traceable and interchangeable thermocouples comprise of a smart USB connector that is molded and sealed. Built-in cold junction compensation is integrated into each USB connector. Each connector maintains the calibration offsets reducing the pre and post calibration time and efforts. 

 

View Thermocouples
PRESSURE & ADDITIONAL PARAMETERS
Other Measuring Parameters for E-Val™ Pro

Unlike most classical thermocouple systems, the E-Val™ Pro has the option of measuring several different parameters other than just temperature. To do this, Ellab designed a Digital Pressure Transducer for pressure measurements and a Sensor Interface which can be fitted with any sensor, thereby covering all available measuring parameters. These sensors are the same ones used in combination with the TrackSense® Pro series of data loggers.

Measuring Parameters:

  • Temperature
  • Pressure
  • Vacuum
  • Relative Humidity
  • Conductivity
  • CO2
     